Research on the calibration method of 10 km/s laser Doppler velocimeter indicating value error
Author:
Affiliation:

Metrology and Testing Center, China Academy of Engineering Physics,Mianyang 621000, China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    对于高速激光多普勒速度测量系统测速误差无法有效校准的问题,本文基于回波模拟法建立示值误差校准方案,通过光频调制与光波长调节模拟测速系统测速过程中携带目标信息的光回波,间接模拟参考速度,进而实现激光多普勒测速系统示值误差的校准。基于自研校准装置搭建校准实验,以测速上限高达10 km/s的激光多普勒测速仪为被校对象,开展示值误差校准实验。通过分析实验结果可知,基于光频调制与波长调节的回波模拟校准方案,可以实现宽速域内测速仪测速误差的校准,由测量准确度分析可知光频调制法的校准不确定度为U=0.26%(k=2),波长调节法的校准不确定度为U=1.5%(k=2)。该指标满足激光多普勒测速仪示值误差的校准需求,为激光多普勒测速仪示值误差的校准提供方法参考。

    Abstract:

    For the problem that the speed measurement error of high-speed laser Doppler velocity measurement system cannot be effectively calibrated, this paper establishes a calibration scheme for the indicated value error based on the echo simulation method, in which the optical frequency modulation and optical wavelength modulation simulate the optical echo that carries the target information in the speed measurement process of the system, and then indirectly simulate the reference speed, so as to realize the calibration of the indicated value error of laser Doppler velocimetry system. Based on the self-developed calibration device, a calibration experiment is built, and a laser Doppler velocimeter with an upper speed limit of up to 10 km/s is used as the calibration object to carry out the calibration experiment of the display value error. By analyzing the experimental results, it can be seen that the echo simulation calibration scheme based on optical frequency modulation and wavelength adjustment can realize the calibration of speed measurement error of the speedometer in a wide speed domain, and the measurement accuracy analysis shows that the calibration uncertainty of the optical frequency modulation method is U=0.26% (k=2), and the calibration uncertainty of the wavelength adjustment method is U=1.5% (k=2). This index meets the calibration demand of laser Doppler velocimeter indicating value error, and provides a method reference for the calibration of laser Doppler velocimeter indicating value error.

    参考文献
    相似文献
    引证文献
引用本文

鲁伟俊,邓文,丁宇洁,樊寅斌,彭希锋.10km/s激光多普勒测速仪示值误差校准方法研究[J].电子测量技术,2025,48(10):8-14

复制
分享
相关视频

文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2025-07-07
  • 出版日期:
文章二维码

重要通知公告

①《电子测量技术》期刊收款账户变更公告