基于超宽带的TDOA相邻单元协同定位技术
DOI:
CSTR:
作者:
作者单位:

1.河北大学 质量技术监督学院 保定 071002; 2.计量仪器与系统国家地方联合工程研究中心 保定 071002; 3.河北省能量计量与安全检测技术重点实验室 保定 071002

作者简介:

通讯作者:

中图分类号:

TN919.72;TP391

基金项目:

国家自然科学基金资助项目(61475041) ;京津冀协同创新共同体建设专项(20540301D);河北省自然科学基金重点项目(F2021201031)资助


Cooperative location technology of TDOA adjacent units based on ultra wideband
Author:
Affiliation:

1. School of Quality and Technical Supervision, Hebei University, Baoding 071002, China; 2. National & Local Joint Engineering Research Center of Metrology Instrument and System, Baoding 071002, China; 3. Hebei Provincial Key Laboratory of Energy Metering and Safety Testing Technology, Baoding 071002, China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    针对目前超宽带TDOA(Time Difference of Arrival)分单元定位系统存在位置多解性、定位精度低等问题,本文提出了相邻单元协同定位方法以及基于改进樽海鞘群算法的协同定位方程解算算法。该方法通过汇总相邻单元内收到同一定位请求信号的全部测量值,使坐标解算不再局限于一个时钟同步单元内,由此也导致参与位置计算的测量值个数与基站布局不确定性增加,Taylor算法无法满足解算要求,进而使用改进樽海鞘群算法代替Taylor算法进行协同定位位置计算,最终求得待测点位置坐标。六基站相邻单元协同定位实验结果表明,各测量点的R95值处于12 cm~15 cm之间,最大残余误差处于16 cm~23cm之间。本文所提方法无需进行多参量判断取舍位置信息,有效解决了位置多解性问题并提高了定位精度。

    Abstract:

    Aiming at the problems of the current TDOA (Time Difference of Arrival) unit positioning system of ultra-wideband, such as multi-solution of position and low positioning accuracy. In this paper, a cooperative positioning method of adjacent units and a cooperative positioning equation solving algorithm based on improved salphus swarm algorithm are proposed. In this method, by summarizing all the measured values of the same positioning request signal received in adjacent units, the coordinate solution is no longer limited to one clock synchronization unit, which also leads to an increase in the uncertainty of the number of measured values and the layout of base stations involved in the position calculation. Taylor algorithm cannot meet the solution requirements, so the improved salphus swarm algorithm is used instead of Taylor algorithm to calculate the cooperative positioning position, and finally the position coordinates of the point to be measured are obtained. The experimental results of the cooperative positioning of adjacent units of six base stations show that the R95 value of each measuring point is between 12 cm and 15 cm, and the maximum residual error is between 16 cm and 23 cm. The method proposed in this paper does not need multi-parameter judgment to choose or reject position information, which effectively solves the problem of multi-solution of position and improves the positioning accuracy.

    参考文献
    相似文献
    引证文献
引用本文

韦子辉,王世昭,叶兴跃,马英杰,李小阳,方立德.基于超宽带的TDOA相邻单元协同定位技术[J].电子测量技术,2022,45(1):77-83

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2024-06-19
  • 出版日期:
文章二维码
×
《电子测量技术》
财务封账不开票通知